Fixed site water sampler in stainless steel housing with thermostatic control for automatic sample extraction according to the vacuum principle. Mains operation 230V/50Hz




Дата канвертавання26.04.2016
Памер18.84 Kb.
Fixed site water sampler in stainless steel housing with thermostatic control for automatic sample extraction according to the vacuum principle. Mains operation 230V/50Hz.
Housing Double-walled stainless steel (material 1.4301) with 40 mm insulation. Consisting of upper part (control unit) and lower part (sample compartment), each with lockable door. Upper door with plexiglass window. Protective top which can be opened for connection and maintenance works.
Thermostatic control Self-contained, automatic cooling and heating of the sample compartment to +4°C independent of the programmable controller, with protection against overheating and automatic defrost cycle.
Control Microprocessor control with 4 MB Flashrom, 32KB RAM (battery-buffered), 32 KB EEprom, 3 digital inputs and 8 digital outputs, 1 configurable analogue input, battery-buffered real-time clock. Operation by means of a waterproof foil keyboard (with keys 0-9, ESC, ENT, cursor) and 4 x 20 character back lit LC-display.
Language Multi-language, selectable
Programming Time display: Hours, minutes, seconds

weekday, day, month, year

Sampling: - Time-related

- Flow-related analogue (0/4-20mA) or digital

Option: flow-proportional

- Event-related or in combination

Interval: 1 min to 15 h

Bottle filling: 1 min to 99 h 59 min

Programs: 6 user programs (for free editing)

Data memory: Logging of sample extraction

and messages.

Optional storage of external data

Interface Optional: RS 232 (internal on CPU board in plug-in version, not led to the outside).
Modem Optional: GSM, AT or ISDN modem to change the programming, for a status fetch, a data fetch or a message transmission as SMS.
PC software Optional: PC software (Win98, NT, XP) for parameterization, status

control, data fetch and direct storage as excel file/text file/PDF

file.
Status messages Optional: Sampling, distributor, program active and collective malfunction

message by means of potential-free make contacts. Can be

extended to up to 10 outputs.

In addition, the desired messages (configurable by means of the software) can be transmitted via modem as SMS or text.


Overvoltage protection All inputs are protected against overvoltage (lightning protection).
Dosing system The following variants are available

-VACUUM SYSTEM

The sample volume is adjusted manually by means of an adjustable tube. Metering vessel made of glass Duran 50, adjustable volume

20 - 350 ml (optional 20 - 500 ml) with motor-driven pinch valve*) .

-VARIABLE VACUUM SYSTEM (flow-proportional sampling)

The dosing volume is automatically adapted to an analogue flow signal

(0/4-20 mA ).

Metering vessel made of glass Duran 50, adjustable volume

25 - 450 ml with motor-driven pinch valve*).

-FLOW-THROUGH SYSTEM

For flow rates from 3 – 20 l/min. Metering vessel made of glass

Duran50, adjustable volume 20 – 250 ml with 2 motor-driven pinch

valves*)



Pneumatic unit Diaphragm pump 12 V/ 4 A, vacuum 7,0 m, pressure 1 bar, motor-driven

valve system*) for change-over pressure/vacuum/aeration, average suction

speed with suction hose ID 12 mm and 5 m suction height > 60 cm/sec

(according to ISO5667).
Distributors The distributors can be retrofitted easily without any tools.
Sample bottles 1 x 25 l PE

1 x 50 l PP

2 x 10 l PE (in combination with 12-step distributor)

4 x 6,5 l PE (in combination with 12-step distributor)

4 x 10 l PE

4 x 14 l PE

12 x 2,9 l PE

12 x 2,9 l + 1 x 5 l (daily composite sample) PE

12 x 2,0 l Glass

24 x 1,0 l PE

24 x 1,0 l Glass

Overall dimensions 1290 (1890*) x 690 x 645 mm (hxwxd)

* with opened top


Weight Approx. 100 kg
Power supply 230 V / 50 Hz, fuse protection 16 A
Optoelectronic coupler

input Analogue, threshold voltage 3,3 V


Power requirement Approx. 350 VA
Ambient temp. -20°C to + 40°C
Standards All devices meet standard ISO 5667

Make: MAXX


Type: SP III
Manufacturer: MAXX Mess- und Probenahmetechnik GmbH

Hechinger Straße 41, D-72414 Rangendingen

Phone +49(0)7471-98481 0, Fax +49(0)7471-98481 44

e-mail: info@maxx-gmbh.com

Internet: www.maxx-gmbh.com

Subject to technical changes. *) Patent No. DE 19726550A1, DE 19726549A1

and VAR unit DE 10008623.3

E-SP-III Fixed site sampler 12/05